FFT view

The FFT view shows the spectra of channels: which frequencies a signal contains, and how strongly. A typical use is the harmonics of a phase current.

FFT view: spectrum of a phase current with its 5th and 7th harmonic

Choosing the spectra

The eyes in the Channel Configuration choose which channels are shown as spectra ("Show/hide spectrum"). This choice belongs to the FFT view: the channels' traces in a time view stay as they are. In a view linked to a time view, the eyes are the only part of the Channel Configuration that can be changed.

Configuration

Configuration of an FFT view

FFT Axes

  • "X axis": "Log" for a logarithmic frequency axis; "Auto", or "Min" and "Max".
  • "Y axis": the scale of the amplitudes, and "Auto" or "Min" and "Max":
Scale Amplitude
Peak Peak value of each frequency, in the unit of the channel
RMS RMS value of each frequency
dB Decibel relative to 1 unit
dBc Decibel relative to the largest amplitude above DC

FFT Processing

  • "Source": where the data comes from, see data source.
  • "Window": "Rectangular", "Hann" (the default), "Flat top" or "Blackman-Harris", see Windows.
  • "Averaging": "Off", "Average" over a number of captures ("Number of captures averaged, 1 ... 100", 8 by default), or "Peak hold". "Reset" starts the averaging anew.
  • "DC removal": removes the mean of the signal before the transform.

Windows

The FFT treats the capture as if it repeated endlessly. A signal with a whole number of periods in the capture fits seamlessly; any other signal has a jump where the capture ends and starts again. Its energy then leaks into the neighbouring frequency lines: a single sine shows a broad skirt, and small lines nearby disappear under it. A window fades the capture in and out, so the jump disappears and the leakage drops. In return, every line becomes wider. The amplitudes are corrected for the window, so a sine on a frequency line shows its amplitude with every window.

Window Width of a line Amplitude between two lines Leakage (highest side lobe) Use for
Rectangular ±1 line Up to 3.9 dB (36 %) low −13 dB Signals with a whole number of periods in the capture, e.g. sampled synchronously; pulses that start and end within the capture
Hann ±2 lines Up to 1.4 dB (15 %) low −31 dB General use: a good compromise between sharp lines and low leakage
Flat top ±5 lines Right to 0.01 dB −93 dB Exact amplitudes of lines at any frequency, e.g. for calibration; lines close together merge
Blackman-Harris ±4 lines Up to 0.8 dB (9 %) low −92 dB Small lines next to large ones, e.g. weak harmonics or the noise floor near the fundamental
  • Width of a line: a sine spreads over this many frequency lines on either side. Two frequencies closer together than about the width cannot be told apart. The line spacing Δf is shown in the status line; a longer capture makes it finer.
  • Amplitude between two lines: a sine whose frequency falls between two frequency lines reads low in the plot and in the cursor read-outs. The Measurements table is not affected: it sums each line over its width, so its amplitudes are right with every window.
  • Leakage: how far the skirt of a line stays below the line itself. The lower, the smaller the lines that remain visible next to a large one.

The harmonics need the fundamental clear of DC: more than 2 (Rectangular), 4 (Hann), 8 (Blackman-Harris) or 10 (Flat top) periods of it in the capture. With fewer, the Measurements table says "Too few periods in the capture for this window".

The plot

The x axis is "Frequency (Hz)"; the y axis is labelled with the scale and unit, e.g. "Peak (A)". The status line below the plot shows the number of captures, the number of samples N, the frequency resolution Δf, the window and the analysed range, e.g. "Samples: 12, N = 2450, Δf = 4.08 Hz, Hann, All".

If the FFT view is linked to a time view whose vertical cursors V1 and V2 are on, only the samples between them are analysed; the status line then shows "V1–V2" and the time range.

If no spectrum can be computed, the plot says why, e.g. "No spectrum shown", "Samples are not equidistant: no spectrum" or "Analysis range too short".

Cursors and markers

  • Vertical cursors V1 and V2 snap to frequency lines. Their read-out shows the frequency f and the amplitude of every spectrum at V1 and V2, and the differences Δ. Left/Right move the cursor moved last by one frequency line.
  • Horizontal cursors H1 and H2 measure amplitudes (y).
  • The multi cursor reads every spectrum out at the frequency the mouse points at ("f = …").
  • Markers mark a frequency line of a spectrum; they follow the new spectra while sampling goes on.

Measurements

The Measurements button shows a table of the visible spectra:

  • Harmonics: the multiples n of the fundamental, with frequency f, amplitude and the amplitude relative to the fundamental (dBc or % of the 1st), followed by THD and THD+N.
  • Peaks: the largest lines of the spectrum, with frequency, amplitude and dBc.

Settings → Measurements... sets the "Number of harmonics or peaks (2..20):" (10 by default); THD sums the harmonics listed.